|
[ A&D ] [ MYWEIGH ] [ HEALTH-O-METER] [ OHAUS ] [ SECA ] [ SCIENTECH ] [ SARTORIUS ] [ SILTEC ] [ TANITA ] |
|
|
|
Warning
- The definitions on this web page are the work of Precision Weighing
Balances and you are NOT granted permission to reproduce or distribute
this page without permission. Another great source to learn more
about weighing definitions is the UK's
National Physical Laboratory
|
The extent to which a given measurement agrees with the standard value for that measurement ¹ The ability of a scale to provide a result that is as close as possible to the actual value. Example, if a known calibration standard weight of 100.000 grams was placed on the Scientech SP250 and the display shows 100.002 grams we could say the accuracy of the balance is 0.002 grams or 2 milligrams. Accuracy tells how close a balance gets to the real value. The accuracy of the scale is very sensitive to the calibration process. It is recommended to perform a calibration at the end user facility. If calibration is well performed, we can usually say the accuracy of the scale should be within +/- one display resolution with most scales.
To determine, check, or rectify the graduations of (any instrument giving quantitative measurements). ¹ Calibration is the comparison between the output of a scale or balance against a standard value. Calibration requires a standard weight and the balance to be set in the "calibration mode."
Calibration technically means to determine the difference between the balance/scale readout and the actual weight on the weighing platform to determine accuracy. Adjustment means to bring a balance/scale into the state of accuracy required for its use. Therefore, 'calibration," actually means "adjustment."
Some instruments contain internal calibration weights that are quite accurate, and can be used by the operator routinely, but we recommend external calibration. The readability of the scale will determine which class calibration mass (Class 1, Class 2, ASTM 6, Class F, etc) will be appropriate for calibrating your balance. Check your operator’s guide since most balances must be calibrated with a specific mass value.
Example: the Ohaus CS200 requires a 200 gram ASTM Class 6 mass. You can not calibrate the CS200 if you have a 100 gram mass. It must be a 200 gram mass. The scale is programmed and the software within the scale is configured only to accept a 200 gram calibration value to store the calibration data within the EPROM on the scale’s motherboard.
There is no certificate of calibration included with any balance we sell since
this can only occur at the place of installation. A calibration certificate
can only be attained at the place of installation for a balance. One can't
calibrate a balance here in Massachusetts and ship it to another destination
with the calibration certificate that would be valid.
The reason for this included:
You can purchase optional
calibration weights for most balances, but the question then becomes do you
needs weights with a "certificate of accuracy", "calibration
report" or "mass value certificate". Read more about calibrations
weights and the huge selection we offer at http://calibration.balances.com/.
Basically local scale
companies are in business just for this reason of calibrating and providing
the documentation requirements. You might want to call a local scale company
to see what they charge. Regardless, calibration certificates is something
neither Precision Weighing Balances nor the scale manufacturer can offer unless
we come on site to install the balance. Calibration is an additional service
and this is always best done exactly at where the balance is in operation.
Another option to attain a certificate of calibration is to do it yourself with
a certified
calibration mass which we sell. You can basically copy the calibration
instructions in the manual and create a QAD (Quality Assurance Directive), QAP
(Quality Assurance Procedure) or SOP. You can also create your own calibration
procedure from scratch like this one here.
People sometimes call us saying a competitor or catalog company offered to ship
their balance with a calibration certificate if they purchase the balance from
them. Companies offering balance(s) with a calibration certificate are
providing false documents unless the calibration is performed on site.
Sure, companies can sell a oscilloscope or power supply with a certificate of
calibration when the power supply is calibrated halfway across the country since
a power supply isn't effected by the different longitude and latitude of magnetic
north like a balance.
Looking for calibration weights? Check
out our huge selection
Calibration error
The difference between what a weight of near the full capacity of the instrument reads on the digital display and its true mass.
The actual or potential ability to perform, yield or withstand. ¹ The largest weight the balance is capable of weighing.
Capacitance Loadcell
The fundamental design of a capacitance loadcell is that of the electrical capacitor. The loadcell contains two closely spaced, parallel, electrically-isolated metallic surface, one of which is essentially a diaphragm capable of slight flexing when pressure is applied. When pressure is applied to the capacitance loadcell a minute change occurs in distance between the plates. The varying gap between the plates creates in effect a variable capacitor. The resulting capacitance is detected send to a linear comparator and amplifier which is then processed by a microprocessor and displayed on the LCD. Dendritics, and many of the Tanita Pocket and Foods Scales use a capacitance loadcell.
Cornerload
Cornerload refers to the ability of an instrument to deliver the same weight reading for a given object anywhere on the weighing pan. (Of course, an instrument that does not perform acceptably with regard to drift and repeatability cannot possibly deliver acceptable cornerload performance.) Test this characteristic using the same test weight that was used to test repeatability. Position the object at various locations on the weighing pan. The reading should be the same, within a few digits, at all positions. A good example and test for cornerload can be found here regarding the Ashiba line of pocket scales.
Cornerload error
Refers to variations in the displayed weight as the object being weighed is moved to various positions on the weighing pan.
Count
The smallest increment of weight which the digital display resolves. Also called "division.".
Digit
The smallest increment of weight that the digital display resolves.
Divisions
The amount of increments a scale offers. The amount of divisions can be determined by taking the scale's capacity divided by the scales readability (the smallest number a scale can display. Example the Ohaus SC4010 features 4,000 divisions. The capacity is 400 grams and the scale's readability or another way to say it is the numbers on the display increase in 0.1 gram intervals. Therefore 400 / 0.1 = 4,000 divisions. Another example would be the Ohaus SC2020 features 20,000 divisions. The capacity for the SC2020 is 200 grams and the scale's readability is 0.01 gram. Therefore 200 / 0.01 = 20,000 divisions. It is the divisions which determines the cost of a scale - not the capacity or readability, but instead the combination of both the capacity and readability to determine the amount of divisions. The more divisions the better the quality of the weighing sensor and larger the A/D converter needed to resolve the analog output from the weigh sensor to a binary number for the digital display.
Drift - a.k.a. Sensitivity
Drift
Drift is a progressive (continuously upward or continuously downward) change in the number displayed on the digital readout. The weight readings does not stabilize, or unstable readings with no weight applied. All analytical balances show some uncertainty. Some do so more than others.
Two environmental factors affect the instrument’s stability dramatically—temperature and static electricity. Temperature control is imperative. This includes both control of the room temperature and maintaining the internal temperature of the instrument. For best stability, maintain the room temperature within two degrees constantly (day and night). Leave the instrument plugged in and turned ON. Static discharge can also be accomplished by putting some ionizing devices around the weighing pan.
Drift may be related to RFI (radio frequency interference). There is not a lot you can do is RFI other than move the balance to a different area where the RFI is less. Leveling of the balance can also result in drift.
External calibration - See detailed information about calibration of balancesFlexible Bearings - Precision components in the measuring cell (force motor) which allow the force coil to move without friction.
Force Motor (a.k.a Electromagnetic Force Compensation) - See a detailed explanation with detailed drawing here. Also call" measuring cell."
![]() |
|
©
Crown Copyright 2004. Reproduced by permission of the Controller of HMSO
|
Hysteresis
- Property of load cells, and other weighing systems dependant on elastic materials,
such as spring balances, resulting in a different indication at the same load,
depending upon the direction of approach to that load, i.e. whether it is approached
by increasing the load or decreasing the load. It is most pronounced at
the mid-point of the weighing range. The lag in response exhibited by a
body in reacting to changes in forces, esp. magnetic forces, affecting it.
The diagram (Fig 1) is from the UK's
National Physical Laboratory and illustrates the concept quite well.
Hysteresis Error - Refers to the condition of repeatedly weighing the same object, but obtaining different readings on the numeric readout.
Instability - the tendency to behave in an unpredictable, changeable, or erratic manner. ¹ Refers to a displayed number which continues to vary randomly or sporadically, rather than progressively. See "drift."
Installation - Learn about selecting the best location of your balance for optimal performance.
Internal resolution - the smallest increment of the A/D converter. It is used by the hardware and software designers. For a scale using a strain gauge design, the ratio between internal and display resolution is about 4:1. For a balances (force motor or SHS design), the ratio can be 10:1. It is possible to use 1:1 ratio, but you will see a lot of unstable readings by changing of 1 increment. Having stable display is the main reason of the ratio. There are other concerns, such as measurement speed and temperature compensations.
A&D offers more flexibility to the counting scale user. At 1/30,000 setup (models HC and HD, the FC is 1/50,000), you can not count very small piece materials, but your counting results will be more stable on the LCD display. It takes a drift of 4 internal counts (at 1/120,000 setup - FC is 1/500,000) to show one count change on the display. Internally, A&D HC and HD counting scales always uses 1/120,000 mode(FC is 1/500,000). This is a benefit/flexibility to customers. In the measurement industry, we care of the most is the "accuracy".
IP
Ratings (Ingress Protection) - The
IP rating system provides a means of classifying the degrees of protection from
solid objects and liquids afforded by electrical equipment and enclosures.
The system is recognized in most European countries and is set out in a number
of British and European standards. These include: Classification of Degrees
of Protection Provided by Enclosures, BS (British Standards) 5490:1977; IEC
(International Electrotechnical Commission) 529:1976.
Specifications for Degrees of Protection of Enclosures of Switchgear and Control
Gear for voltages up to and including 1000 VAC and 1200 VDC, BS 5420:1977; and
IEC 144:1963.
First number Protection against solid objects
0 - no protection
1 - protected against solid objects up to 50 mm (e.g. accidental touch by hands)
5 - protected against dust-limited to ingress (no harmful deposits)
6 - totally protected against dust
Second number Protection against liquids
0 - no protection
1 - protected against vertically falling drops of water (e.g. condensation)
2 - protected against direct sprays of water up to 15° from the vertical
4 - protected against water sprayed from all directionslimited ingress
permitted
6 - protected against strong jets of waterlimited ingress permitted (e.g.
for use on ship decks)
7 - protected against the effects of immersion between 15 cm and 1 m
8 - protected
against the effects of extended periods of immersion under pressure
IP-43 Rating - The first digit designation "4" means protection against "solid objects equal or greater than 1mm (0.04") in diameter". The second digit "3" means protection against "sprayed water" ; "Spray water falling perpendicularly at an angle of up to 60-deg may not have any damaging effects".
IP-54
Rating - The "5" means
"Dust Protected. Ingress of dust is not totally prevented, but dust does
not enter in sufficient quantity to interfere with satisfactory operation of
the equipment. The "4" means "Protected against splashing water.
Water splashed against the enclosure from any direction shall have no harmful
effect."
IP-66
Rating - A strong water
jet directed at the scale from any direction must not have any harmful effects.
A jet nozzle with an inside diameter of 0.49 inches splashes a volume flow of
approximately 26.4 gallons per minute, from a distance of approximately 8.2
- 9.8 Feet from all sides onto the scale. The test time is 3 minutes.
IP-67
Rating - Temporary Submersion.
The device is placed in water at a depth of 3.28 feet for 30 minutes
IP-68
Rating - Long Term Submersion
IP-69K Rating - A strong water jet directed at the sensor from 4 directions must not have any harmful effects. A jet nozzle at 0°, 30°, 60° and 90° to the scale on a rotating table at 176° + 8°F, 4-6 inches away at 1250-1500psi. The test time is 2 minutes.
Light Emitting Diode (LED) - A type of solid-state numeric readout device. Often characterized by red or orange-red numbers.
Linearity refers to the quality of delivering identical sensitivity throughout the weighing capacity of a balance or scale. Test this characteristic by weighing two stable objects separately, each of approximately one half the weighing capacity. The sum of the two readings should equal the reading obtained when both objects are weighed together.
Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes deviation between actual and displayed weights within the balance's weighing range.
Perhaps the most obvious test of a high precision scale would be to place a weight of accurately known value on the weighing pan, and observe the numerical result. But there is a better test, nearly as simple, that better reflects the measurement accuracy. This is called the linearity test.
The linearity test measures the ability of an instrument to have consistent sensitivity throughout the weighing range. The test requires several nominally equal weights, each a fraction of the weighing capacity. The group together should approximate the weighing range of the instrument. For example, a 160 gram capacity analytical balance might be tested with three 50 gram weights.
Static electricity will cause erratic readings. Instruments should be operated on a static dissipating surface (antistatic mat). Operators should stand on antistatic floor covering. Avoid the use of plastic containers for items being weighed. Never replace broken glass doors on instruments with plastic ones. Maintain humidity at 65% or more. Eliminate sources of floor vibration and air currents. On analytical balances with glass doors, be sure the doors close fully.
Liquid Crystal Display (LCD) - A numeric readout device, often characterized by black numerals on a silver background.
Mass
tolerances - weight
classifications & applications:
ASTM Class 1: Can be used as a reference standard in calibrating other weights
and is appropriate for calibrating high precision analytical balances with a
readability as low as 0.1mg to 0.01mg.
ASTM Class 2: Appropriate for calibrating high-precision top loading balances
with a readability as low as 0.01g to 0.001g.
ASTM Class 3: Appropriate for calibrating balances with moderate precision,
with a readability as low as 0.1g to 0.01g.
ASTM Class 4 -For calibration of semi-analytical balances and for student use.
NIST Class F - primarily used to test commercial weighing devices by state and
local weights and measures officials, device installers and service technicians.
Class F weights may be used to test most accuracy class III scales, all scales
of class III L or IIII, and scales not marked with a class designation. Calibrated
according to NIST Handbook 105-1
OIML Class E1 - Used as primary reference standards for calibrating other reference
standards and weights.
OIML Class E2 - Can be used as a reference standard in calibrating other weights
and is appropriate for calibrating high precision analytical balances with a
readability as low as 0.1mg to 0.01mg.
OIML Class F1 - Appropriate for calibrating high-precision toploading balances
with a readability as low as 0.01g to 0.001g.
OIML Class F2 - For calibration of semi-analytical balances and for student
use.
OIML Class M1, M2, M3 - Economical Weights for general laboratory, industrial,
commercial, technical and educational use. Typically fabricated from cast
iron or brass. Class M2 commonly used for student brass weights.
Measuring Cell - (Also known as a force motor.) is the part of the instrument which "senses" weight, and converts it to an electrical signal. Similar to "mechanical system" or "force motor."
Min Weight - typically used in the specifications of counting scales. The small piece weight required in a counting mode. At the beginning of any counting process, the scale's software needs to teach the scale what is the unit piece weight. The scale uses the information to count the unknown weight.
A&D offers more flexibility to the counting scale user. At 1/30,000 setup (models HC and HD, the FC is 1/50,000), you can not count very small piece materials, but your counting results will be more stable on the LCD display. It takes a drift of 4 internal counts (at 1/120,000 setup - FC is 1/500,000) to show one count change on the display. Internally, A&D HC and HD counting scales always uses 1/120,000 mode(FC is 1/500,000). This is a benefit/flexibility to customers. In the measurement industry, we care of the most is the "accuracy".
Moisture Balance heating elements:
NTEP / Type Approved / Legal for Trade - The National Type Evaluation Program (NTEP) evaluates commercial measuring devices and issues Certificates of Conformance to those devices that meet national and/or international standards. In this way, consumers are protected and can be confident that they are charged fairly for goods or services rendered.
Type evaluation is the process whereby weighing and measuring devices are examined by government officials to determine if the design and performance comply with the Weights and Measures requirements of the country in which the manufacturer wishes to market the device for commercial applications.
The purpose of type evaluation is to examine the design, features, operating characteristics, and performance of weighing and measuring devices for compliance with Weights and Measures requirements. Such tests are performed on sample devices under laboratory conditions. Read more about NTEP approved / legal for trade scales here. To see all the Legal for Trade, Class III scales we offer click here.
Precision - The extent to which a given set of measurements of the same sample agree with their mean. ¹ Amount of agreement between repeated measurements of the same quantity. Also know as repeatability. A scale can be extremely precise, but not necessarily be accurate. Example, two balances were evaluated for precision. Both balances: Balance "A" and Balance "B" offers 200g x 0.001g. A 100.000 gram ASTM Class 1 test mass was place on each balance 70 times.
Balance "A" displayed 103.005 grams 68 times and 103.004 grams 2 times.
Balance "B" displayed 100.000 grams 10 times, 100.001 grams 11 times, 100.002 grams 9 times, 100.003 grams 17 times, 100.004 grams 17 times, 99.999 grams 4 times, 99.998 grams 10 times and 99.996 grams 2 times.
Conclusion: Balance "A" is more precise even though the balance measured a 100.000 test mass as 103.005. While Balance "A" is more precise Balance "B" is more accurate since it measured the 100.000 test mass more to the actual mass value.
Readability - Smallest division at which the balance’s LCD increments.
Examples:Ohaus CS2000 features a 2000 gram weighing capacity and increment in 1 gram increments (2000g x 1g.) Therefore, the readability is 1 gram. The LCD will increment 1 g, 2 g, 3 g, 4 g, .... 1999 g, 2000 g. You will never see 0.1 g or 0.5 g with the CS2000. The scale manufacturer also defines the CS2000 scale with an accuracy of +/- 3 grams, but a readability of 1 gram. Therefore the scale increments in 1 gram intervals but it is on accurate to +/- 3 grams.
Tanita KD200-110 increments in 1 gram intervals. This means when weighing item(s) on the weighing platform from 0 to 1000 grams the LCD will increment from 0 to 1 g, 2 g, 3 g, ... 999 g, 1000 g. Again you will never see the LCD show 0.1g or 0.5 grams. The display will ALWAYS BE IN WHOLE GRAM increments. Therefore, if you need 0.1 accuracy you need to look at A&D EK1200G (1200g x 0.1g) since they don't make a scale that is 1000g x 0.1g..
Tanita 1140 is a dual range scale. This means when weighing item(s) on the weighing platform from 0 to 500 grams the LCD will increment from 0 to 1 g, 2 g, 3 g, ... 499 g, 500 g. Again you will never see the LCD show 0.1g or 0.5 grams. The display will ALWAYS BE IN WHOLE GRAM increments. When the weight on the platform exceed 500 grams (501 - 1000g) the LCD will increment in 2 gram intervals automatically. You have no control of this since the manufacturer has programmed the scale to operate in 2 gram intervals when anything is placed on the platform over 500 gram. Therefore, the display will show 502 g, 504 g, 506, up to 996 g, 998 g, 1000 grams.
The Tanita 1479(grams only version), Tanita 1471(grams & ounce version), Tanita 1477 (grams & pennyweight), Tangent KP102(gram only) and many other pocket scales operate with a dual range except they read at 0.2 grams. This means when weighing item(s) on the weighing platform from 0 to 50 grams the LCD will increment from 0.0 to 0.1 g, 0.2 g, 0.3 g, ... 49.9 g, 50.0 g. Again you will never see the LCD show anything less than 0.1g when purchasing a pocket scale. When the weight on the platform exceeds 50.0 grams (50.1 - 100.0g) the LCD will increment in 0.2 gram intervals automatically. You have no control of this since the manufacturer has programmed the scale to operate in 0.2 gram intervals when anything is placed on the platform over 50.0 gram. Therefore, the display will show 50.2 g, 50.4 g, 50.6, up to 99.6 g, 99.8 g, 100.0 grams.
Reproducible - Refers to the ability of an instrument to return the same numeric result with repeated application of the same weight. See " hysteresis."
Resolution - The smallest increment of weight which the numeric display can indicate. Also referred to as 'display resolution".
Repeatability refers to an instrument’s
ability to consistently deliver the same weight reading for a given object,
and to return to a zero reading after each weighing cycle. Test this by
repeatedly weighing the same object. The best test object is a weight
intended for that purpose. It should match the weighing capacity of the
instrument. (Do not test a 200 gram capacity instrument with a test weight
less than 100 grams) When a test weight is not available, an alternative
object that is solid, non-porous, dirt free, non-magnetic, and non-static retaining
can be used. Repeatability is sometimes referred to as "Standard
Deviation" of a set of similar weight readings.
Sensitivity
Drift
Here are two explanations
for sensitivity to drift:
The sensitivity drift is how a change in temperature can affect the performance
of the balance. Basically it states what the user can expect to see from
the balance when the temperature changes. This phenomenon is more relevant on
micro balances which offer resolutions
up to 0.001mg. The change in temperature is measured by a thermistor in
the weighing cell which sends the value to the A/D converter module.
As
an example, the A&D GF-600
which has a specification for sensitivity to drift at ±
2ppm /°C. If there were 30g on
the balance & there was an increase in the temperature of 10°C
the weight change would be 0.0006g [weight in grams
x sensitivity
to drift x
temp change in °C so 30
x (0.000002 x 10)].
With the resolution of the balance being at 1mg (0.001g) you would most likely
only see a change of 1 division. Obviously the balance would show more
of this effect as the change in temperature increases.
Here is a second explanations
for sensitivity to drift:
All electronic balances are susceptible to a weight reading
drifting due to environmental temperature changes. The measure of this
drift is called sensitivity drift.
Sensitivity drift relates this temperature affect at a given weight. For
example, for a 2.1g x 0.1ug Sartorius
SE2 micro balance with a sensitivity drift spec of ± 1ppm per degree
C and a 1 gram weight on the pan, the balance will drift 1 ug (0.000001g) for
every degree of change in temperature. So if your micro balance is in
a separate room with no temperature control and 2 or 3 person walk into the
room resulting in a temperature increases by 2°C sensitivity of drift becomes
an issue. If the test was to determine particles captured on a 2 gram
filter where the filter is weighed, then the filter is placed into the exhaust
air stream of a piece of equipment being measure for emissions and the filter
is now weighed again the delta might be 10ug. The problem is if the temperature
was raised by 2°C then 4 ug [weight
in grams x sensitivity
to drift x
temp change in °C so 2
x (0.000001 x 2)]
is the result of sensitivity to drift
and 6ug is the weight of particles captured in the filter. Here sensitivity
to drift dramatically effects the weighing results.
Every time you recalibrate a balance, this drift starts over at zero.
This is why Sartorius puts their ISO Cal technology into a balance and why some
metrology handbooks recommend recalibrating a balance at different times of
the day.
As you can tell, for anything other than micro balances, it really does not
matter that much as the drift spec is lower than the repeatability and linearity
specs of most of the models offered.
Many manufacturer's do not have a sensitivity drift spec for any of their models
including the 0.1mg models since sensitivity to drift only effects balances
with readability to 0.01mg, 1ug or 0.1ug.
If a customer is really concerned about sensitivity drift, the best thing for
them to do is recalibate the balance on a continual basis.
Span calibration utilizes two calibration points, one at zero and a choice of either half capacity or full capacity.
![]() |
![]() |
|
Commonly called a loadcell. Consists of an aluminum beam with 4 strain gauges bonded at the hinge areas. The strain gauge is packaged as a film and when the strain gauge is bent the resistance value changes, similar to a potentiometer. The 4 gauges are wired to form a wheatstone bridge. When a load is placed on the beam, it bends at the flexures. The bending changes the resistance value of the strain gauges (normally 350 ohms at rest) and the resulting output from the wheatstone bridge is proportional to the load.
With the basic understanding of electronics including "Ohm’s Law" in which Voltage = Current x Resistance and the understanding that current is the same in a series circuit you will be able to understand the functionality of a load cell.
Two points of the wheatstone bridge are connected to an exciter voltage (from the battery or AC adapter) and an output analog voltage feed to a A/D Converter. The output voltage being fed in the A/D varies in proportion to the load applied to the platform of the scale. This occurs since the weighing platform of a scale is connected to the end of the load cell via a post. The applied force is transferred from the platform, through the post and onto the aluminum beam. Since the aluminum beam is milled out in the shape of a dog bone, the force applied results in a deformation of the beam. When the beam bends the strain gauges bend resulting in their resistance value to change. Since the current is the same in a series current and the resistance changes as the strain gauges bends, the voltage changes in proportion to the load applied to the platform.
The analog voltage is converted by the A/D into a digital signal which is processed by a microprocessor and the microprocessor outputs the appropriate control signals to illuminate the corresponding segments of the LCD to display the correct numerical number for a user to read from the LCD. In between the steps there are some filters, but basically there really isn’t much to a scale.
The problem with scales incorporating a load cell is that an excessive load can permanently bend the load cell. If the aluminum beam is permanently bent the scale will not work. This is NOT covered under the scale manufacturer’s warranty. In most situations the cost of replacing the load cell cost almost as much a replacing the scale, because of the ‘deep discount prices" we offer. For example Ohaus charges approximately $90.00 to replace the load cell on the SP401 while we sell a brand new Ohaus SP401 for $142.00.
Therefore, great care needs to be taken when using a scale with a load cell. The person needs to have an idea how much an object weighs prior to placing it on the platform. Otherwise, you can destroy the scale.
The best way I describe this is if you go out and buy a brand new bicycle. Everything works great and you cruise down the street and hit a small pot hole. The bicycle’s rim doesn’t get bent yet, but you have stressed the metal of the rim. Now you decide to jump a few curbs and the rim is still okay. You feel stupid today, and decide to hit that pothole again. Now the rim is bent. You can’t go back to the bicycle dealer and say the bike is junk since the rim in bent. The rim got bent due to abuse and this is not covered under the manufacturer’s warranty. Well, the same works with scale that incorporates a loadcell. Many people don’t have an idea what something weighs prior to putting the object on the platform but you really need to get acquainted with what objects weigh prior to placing the object on the platform. The manufacturers have taken precautions by installing a down stop which restricts the downward movement of the loadcell but prudence still must be followed.
Switches
Membrane Switches or displays
that use tactile membrane switch panel are relible but they don't work in a
vacuum. These use a foil contact point.
Micro Switches are switches that make contact when the micro switch is pushed.
Typically micro switches cost more and are in the more expensive balances like
the Acculab
ALC Series or Sartorius Extend Series.
Act of removing a known weight of an object, usually the weighing container, to zero a scale. Taring allows you to display the weight of the material on the scale's LCD with the weight of the material only and not the material and container. Most balances allow taring to 100% of the weighing capacity.
Tare by subtraction means that you can keep on using the tare button providing the total mass on the platform does not exceed the weighing capacity of the scale. Therefore, if you had the Tanita KD200-510 (5000g x 5g) and you put a pot on the scale that weighed 1000 grams and pressed the tare button the scale would display 0.0 and you would now have 4000 grams weighing capacity left (5000 - 1000 = 4000 g.) Then you put 500 gram olive oil in the pot and press the tare button the scale would display 0.0 and you would now have 3500 grams weighing capacity left (5000 - 1000 - 500 = 3500 g.) etc, etc.
Digital scales are electronic devices. All electronic devices contain electronic components that have temperature coefficient. An example of this would be a resistor. If you measure a resistor with an Ohm meter and it measures 10 ohms at 0 degrees F and you then put that resistor in an environment chamber and bring the temperature up to 100 degrees F the resistance's value now could be 10.2 ohms. Since the resistance value has increased this is an example of a discrete component(a resistor) having a "positive temperature coefficient". Well, enough with electronics 101.
How does operating a scale above the manufacturer's operational temperature effect you? The temperature operational range is stated since the manufacturer has tested and confirmed that his scale will have an accuracy (scale manufacturers' use the word "LINEARITY") of +/- however many grams provided you operate the scale within the stated temperature. If you go outside of this temperature the scale may be off by a division (in the case of the Ohaus SC6010 <600 g x 0.1g> a division is 0.1 gram so that means it maybe off by +/-0.2 grams instead of the stated +/- 0.1 gram). Like all electronic devices it is not a good ideal to operate them all day at excessive temperatures. Will they operate?, Yes. Consider your computer, this also has a temperature operating range. Can you run your computer in an environment that is a 100 degrees?, Yes. Do you want to do this all the time?, No because it stress the components.
General Product Features
Power Up Test: When the scale is turned on, all display segments will appear for approximately 3 seconds before resetting to zero.
Stable Reading Indication: During weighing, a segment of the display activated once a stable reading has been reached.
Overload: If the applied load exceeds the capacity of the scale, an "E"' will appear on the display and the load should be removed immediately. The scale will return to normal operation. Excessive overloading the scale can destroy the load cell and this is not covered under warranty.
Negative Value: When a load is removed from the scale, any tared value will be displayed as a negative number. To return to normal operation, the tared value can be canceled by pressing the tare button.
Zero Function: Values can progressively be added to a sample. By pressing the tare key, the scale display returns to zero and an indication appears at the upper left corner of the display.
Auto Shut-Off: To extend battery life, the scale will automatically turn off after approximately two to five minutes(depending on scale manufacture) if no active weighing is occurring.
Off: Pressing this key turns the scale off.
Conversions
1 grain = 0.0648 grams
1 gram = 15.4324 grains
1 gram = 0.643 pennyweights
1 gram = 0.03215 troy ounces
1.55517 grams = 1 pennyweight
28.3495 grams = 1 avoirdupois ounce
31.10348 grams = 1 troy ounce
1 kilogram = 32.15076 troy ounces
1 pennyweight = 24 grains
1 pennyweight = 0.05 troy ounces
20 pennyweights = 480 grains
20 pennyweights = 1 troy ounce
14.583 troy ounces = 1 avoirdupois pound
1 troy ounce = 1.09714 avoirdupois ounce
|
Warning
- The definitions on this web page are the work of Precision Weighing
Balances and you are NOT granted permission to reproduce or distribute
this page without permission. Another great source to learn
more about weighing definitions is the UK's
National Physical Laboratory
|
|
[Hot Buys] [Analytical Balances] [Baby Scales] [Counting Scales] [Industrial Platform & Bench Scales] [Mechanical Balances] [Moisture Balances] [Pocket Scales] [Portable Balances] [Precision Toploader Balances] [Price Computing Scales] [Bodyfat Scales] [Scales under $100] [Contact Us] [Return Policy] [Home] Have Questions? Call 978-521-7095 or e-mail us your questions and we will assist you in your selection. Copyright© 2010 Precision Weighing Balances All rights reserved. http://www.balances.com, balances.net, scaleman.com, scalenet.com, weigh.net, weigh.org, analytical-balances.com, lab-balances.com, moisture-balances.com, & precision-balances.com are all registered domains of Precision Weighing Balances. This site is owned and operated by Precision Weighing Balances an independent reseller of balances, scales and weighing accessories. This site is not sponsored or affiliated with any of the manufacturers and should not be confused with the actual manufacturer's web site. |